Three-dimensional assessment of susceptibility-induced signal voids caused by active cardiac implants in the setting of 3.0T CMR

Theresa Reiter, Ingo Weiss, O. Weber, Wolfgang R. Bauer

Abstract

All measurements were obtained at a clinical 3.0T scanner using an in-house designed cubic phantom and optimized sequences for artifact evaluation (3D gradient echo sequence, multi-slice 2D turbo spin echo sequence). Reference sequences according to the American Society for Testing and Materials (ASTM) were additionally applied. Four representative active cardiac devices and a generic setup were analyzed regarding volume and shape of the signal void. For analysis, a threshold operation was applied to the grey value profile of each data set.