Phase Error Correction in Time-Averaged 3D Phase Contrast Magnetic Resonance Imaging of the Cerebral Vasculature Correcting Phase Errors in PC-MRI of the Cerebral Vasculature

M. Ethan MacDonald, Nils D. Forkert, G. Bruce Pike, Richard Frayne, Heye Zhang

Abstract

Phase contrast (PC) magnetic resonance (MR) imaging is a powerful method for measuring blood flow velocity in human subjects and can provide key hemodynamic information about the vasculature. However, concern has been raised about measurement accuracy , in particular, the impact of phase error from the application of magnetic gradient fields causing a spatially varying offset in the phase across the image . This bias can limit the accuracy of quantitative PC-MR measurements. For some regions, like the heart, this phase error has been reported to be as large as 10% to 25% . There are two mechan